← Back to Products
Patient Monitors
Patient Monitor (BM-K12)
Reliable medical equipment engineered for clinical performance and hospital use.

Product Details
Specifications
ECG
- Input dynamic range: ±(0.5mVp~5mVp)
- Differential input impedance: ≥10MΩ
- Bandwidth: 0.05~150Hz (Diagnostic)
- 0.5~40Hz (Monitoring)
- 1~20Hz (Operation)
- CMRR: ≥90dB (Diagnostic)
- ≥105dB (Monitoring & Operation)
- Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto
- Sweeping speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
- HR measuring range: 15~350bpm
- HR accuracy: ±1% or ±2bpm, whichever is greater
- Pacemaker pulse detection and rejection function
RESP
- Measuring range: 0~120rpm
- Measuring accuracy: ±5% or ±2 rpm, whichever is greater
TEMP
- Measuring range: 21.0~50.0°C
- Measuring accuracy: ±0.2°Cfrom 25~45°C
NIBP
- Technique: Oscillometric method
- Typical measurement time: <30 seconds (adult cuff)
- NIBP measuring range SYS: 40~275mmHg (Adult), 40~200mmHg (Pediatric), 40~135mmHg (Neonate)
- NIBP measuring range DIA: 10~210mmHg (Adult), 10~150mmHg (Pediatric), 10~95mmHg (Neonate)
- NIBP measuring range MAP: 20~230mmHg (Adult), 20~165mmHg (Pediatric), 20~110mmHg (Neonate)
- NIBP measuring accuracy (mean difference): ±5mmHg
- NIBP measuring accuracy (standard deviation): 8mmHg
- NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode
- Auto measuring intervals: 1-480min
SpO2
- Technique: Dual-wavelength optical method
- Measuring range: 0%~100%
- Measuring accuracy: Arms is not greater than 2% for
- SpO2 range 70~100%.
- PR measuring range: 30~250bpm
- PR measuring accuracy: ±2bpm or ±2%, whichever is greater
- Low perfusion performance: As low as 0.3%
CO2
- Technique: Infrared optical method
- Sampling mode: Sidestream or Mainstream
- Measuring range: 0~150mmHg
- Measuring accuracy: 0~40mmHg ±2mmHg, 41~70mmHg ±5% of reading, 71~100mmHg ±8% of reading, 101~150mmHg ±10% of reading
- Flow rate: 50ml/min ±10 ml/min (Sidestream)
Cerebral State Monitoring (CSM)
- EEG sensitivity: ±400µV
- Noise level: <2µVp-p, <0.4µV rms (1~250Hz)
- CMRR: >140dB
- Input impedance: >50Mohm
- CSI and update: 0-100. filter: 6-42Hz, 1 sec. update
- EMG%: 0-100 (logarithmic) filter: 75-85 Hz, 1 sec. update.
- BS%: 0-100. filter: 2-42 Hz, 1 sec. update
IBP
- Technique: Strain gauge transducer
- Input sensitivity: 5μV/V/mmHg
- Measuring range: -50~300mmHg
- Measuring accuracy: ±2% or ±4mmHg, whichever is greater
- Measuring positions: ART, RAP, PA, LAP, CVP, ICP, AUXP1, AUXP2
- Calibration: zero calibrating
Cardiac Output (C.O.)
- Blood temperature measuring: range: 23-43°C, accuracy: ±0.5°C
- Injecta temperature measuring: range: 0-20°C, accuracy: ±0.5°C
- Measuring range: 0.2~20 L/min
- Measuring accuracy: ±0.2 L/min or ±10%, whichever is greater
Other
- Power supply: AC 100V-240V, 50/60Hz, 60VA
- Built-in lithium battery: 11.1V/4400mAh
- Display: 12.1 inch TFT display
- Alarming method: 3 levels audible-visible alarm
- Networking: Ethernet
Standard configuration
ECG, Respiration, SpO2, PR, NIBP, Temperature
Options
2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG, Cardiac Output Cerebral State Monitoring, Central Monitor Station, Touch Screen